VACUUBRAND MV 2 NT VARIO User Manual

Page 93

Advertising
background image

page 93 of 106

Loosen the union nuts (N) of the hose con-

nections at the pump heads with an open-

ended wrench (w/f 15).

Turn the fittings (Q) with an open-ended

wrench (w/f 20) to detach the hoses from

the pump heads.

+

Pull off the one curved connecting hose (R)

at the outlet side of the pump directly at the

pump head.

Loosen the fittings (Q) at the pump heads

with an open-ended wrench (w/f 17).

Disassemble the housing cover (A) to check

the valves (C).

Unscrew four Allen screws with a 5mm wide

Allen key. Remove the housing cover (A) to-

gether with head cover (D), valves (C) and

O-rings (B).

+

Never use a pointed or sharp-edged tool to remove parts (e.g., screw-

driver). We recommend to use a rubber mallet or compressed air (to

be blown carefully into port).

Remove the head cover (D) carefully from the housing cover (A). Note

position and alignment of valves (C). Remove the valves.

+

Replace valves or O-rings if necessary.

+

Use petroleum ether or other industrial solvent to remove deposits. Do

not inhale vapors.