Replacing the valve at the distributor (outlet side)

Unscrew the two countersunk head screws

(M) at the cover plate (L) with an Phillips

screw driver. Remove cover plate.

Loosen the union nut of the curved connecting hose (R) running di-

rectly to the cover plate (L) of the distributor (P) with an open-ended

wrench (w/f) 15. Loosen only the union nut at the pump head, not the

one at the distributor.

+

Pull off the curved connecting hose at the pump head.

M

Note the position and orientation of the valve

(B) and remove.

+

Check valve for damage and replace if nec-

essary. Make sure that the valve is correctly

seated.

Reassemble cover plate (L) and secure with

countersunk screws (M).

B

Assembling the connecting hoses

Slip the connecting hoses (O) onto the hose

connectors by turning the fittings (Q) with

an open-ended wrench w/f 20.

+

Mount the one curved connecting hose (R)

at the outlet side of the pump directly at the

pump head.

Tighten the fittings (Q) at the pump heads

with an open-ended wrench w/f 17.