VACUUBRAND MV 2 NT VARIO User Manual

Page 82

Advertising
background image

page 82 of 106

Fault

Possible cause

Remedy

❑ No digital pressure

reading.

➨ Pressure transducer

defective?

✔ Contact local distributor.

❑ Digital pressure

reading is flashing,

one blip*.

➨ Overpressure at the

pressure transducer,

pressure > 795 Torr

(1060 mbar)?

✔ Release pressure imme-

diately (risk of bursting).

❑ Warning triangle

and black valve

symbol are flash-

ing, two blips*.

➨ External venting valve

removed or defective?

✔ Connect valve or replace

with a new one or recon-

figure without valve.

❑ Warning triangle

and valve symbol

are flashing, three

blips*.

➨ NT VARIO / VARIO-B

pump and in-line valve

connected?

✔ Disconnect in-line valve;

switch controller off/on to

reconfigure.

❑ Warning triangle

and coolant valve

symbol are flash-

ing, four blips*.

➨ Coolant valve removed

or defective?

✔ Check connection cable

of the valve; or use new

valve or reconfigure with-

out valve.

❑ Warning triangle

and pump symbol

are flashing, six

blips*.

➨ NT VARIO / VARIO-

B pump and VMS**

(Vacuum Management

System) connected?

✔ Remove VMS. Restart

controller.

➨ Fault at the NT VARIO /

VARIO-B pump?

✔ Check pump, restart con-

troller.

❑ Clock symbol is

flashing.

➨ Preselected process

time is over?

✔ Confirm by pressing key

START/STOP.

❑ Venting valve does

not respond, valve

symbol is dis-

played.

➨ Venting valve contami-

nated?

✔ Clean valve.

❑ ”Vac control” func-

tion: Control stops,

”arrow up” is flash-

ing.

➨ Preset maximum pres-

sure exceeded?

✔ Confirm by pressing key

START/STOP. Change

maximum pressure value

if necessary.

❑ ”Pump down” func-

tion: Control stops,

”arrow down” is

flashing.

➨ Pressure below preset

minimum pressure?

✔ Confirm by pressing key

START/STOP. Change

minimum pressure value

if necessary.

❑ Controller does

not respond when

pressing keys (ex-

cept ON/OFF).

PC symbol is dis-

played.

➨ Controller in remote

mode?

✔ Control CVC 3000 via in-

terface or switch off re-

mote mode.

Advertising
This manual is related to the following products: