When assembling, ensure

vacuum-tightness. After as-

sembly, check the whole system for leaks.

Secure hose connections at the pump appropriately, e.g.,

with hose clamps, to protect against accidental detach-

ment.

The VACUUBRAND controller CVC 3000 can only be

operated with components compatible with the VACU-

UBRAND VACUU•BUS system, (see “Accessories”, pg.

80).

The vacuum controller CVC 3000 controls VACUUBRAND

diaphragm pumps NT VARIO and VARIO-B, and optional

coolant and venting valves. Connected components (e.g.,

venting valve, external pressure transducer 3000 series)

are automatically identified and configurated. Identical

components must be configurated beforehand; informa-

tion upon request.

Connect the controller CVC 3000 via the VACUU•BUS ca-

ble of the NT VARIO or VARIO-B pump to the diaphragm

pump.

Do not use more than one controller within the same

VACUU•BUS system.

Several controllers in the same VACUU•BUS system

will interfere with each other and result in error mes-

sages of the connected components (pumps, valves).

VSK 3000

The VSK 3000 is equipped with a capacitive pressure

transducer with ceramic diaphragm. It measures the ac-

tual pressure independently of the gas type, and with ref-

erence to the vacuum, i.e., absolute.

Connect the VSK 3000 pressure transducer via the

VACUU•BUS cable to the CVC 3000 controller.

Attention: Do not apply off-axis forces when assem-

bling or removing plug connections! Observe correct ori-

entation of the plug. To connect further components use

VACUU•BUS Y-adapters and extension cables.