Position new diaphragm (F) between dia-

phragm clamping disc with square head

screw (E) and diaphragm support disc (G).

+

Note: This is a double diaphragm consist-

ing of two single diaphragms! Put the two

diaphragms together with the printed sides

outwards.

+

Make sure that the square head screw of the diaphragm clamping disc

is correctly seated in the guide hole of the diaphragm support disc.

Lift the diaphragm at the side. Place the diaphragm carefully together

with diaphragm clamping disc and diaphragm support disc in the dia-

phragm key.

+

Avoid damage of the diaphragm: Do not excessively bend or crease

the diaphragm.

+

Too few washers: The pump will not attain vacuum specification. Too

many washers: Diaphragm clamping disc will hit head cover, causing

noisy operation and possibly causing the pump to seize up.

+

If the old diaphragm is difficult to separate from the diaphragm support

disc, immerse assembly in naphtha or petroleum ether. Do not inhale

vapors!

E

G

F

+

Assemble the original number of washers

(H) between diaphragm support disc (G)

and connecting rod (I).

Screw diaphragm clamping disc (E), dia-

phragm (F), diaphragm support disc (G),

and washers (H) to connecting rod (I).

Optimum torque for the diaphragm support disc: 4.4 ft

.

lb

f

(6 Nm), it is

recommended to use a torque wrench. Attach torque wrench to dia-

phragm key (hexagonal bolt 6 mm wide).

Note: Never use the diaphragm key with any additional tools like tongs

or Allen keys without appropriate torque limitation.