Operation Theory

53

The trigger signal asserts when an analog trigger condition is met.
There are five analog trigger conditions in DAQ-/DAQe-/PXI-2502/
2501 card. The DAQ-/DAQe-/PXI-2502/2501 card uses two
threshold voltages: Low_Threshold and High_Threshold to com-
pose five different trigger conditions. You can configure the trigger
conditions easily via software.

Below-Low Analog Trigger Condition

Figure 4-26 shows the below-low analog trigger condition, the
trigger signal asserts when the input analog signal is lower than
the Low_Threshold voltage. High_Threshold setting is not used
in this trigger condition.

Figure 4-26: Below-Low Analog Trigger Condition

Trigger Level digital setting Trigger voltage

0xFF

9.92V

0xFE

9.84V

---

---

0x81

0.08V

0x80

0

0x7F

-0.08V

---

---

0x01

-9.92V

0x00

-10V

Table 4-8: Ideal Transfer Characteristic of Analog Trigger SRC1