Measuring the average loudness of identified clips – Dolby Laboratories DP570 User Manual

Page 43

Advertising
background image

Using the DP570 Features

Measuring and Setting the Dialogue Level

Dolby

®

DP570 Multichannel Audio Tool User’s Manual

31

characters with different vocal qualities. In programs without dialogue, such as a music 
concert, sections of average level need to be identified.

With representative audio clips identified, you can use the DP570 to set the dialogue level 
parameter. You can use either the software interface or the front‐panel buttons; for this 
operation we describe using the software interface first.

4.7.2 Measuring the Average Loudness of Identified Clips

The measurement feature complies fully with IEC 60804 (2000‐10): Integrating‐Averaging 
Sound Level Meters (Leq(A))
 and includes the addition of two distinct measurement modes 
relating to the measurement period itself: infinite and short‐term. This section describes the 
behavior of both modes.

IMPORTANT NOTE REGARDING MEASUREMENT ACCURACY:

Leq(A)‐based measurements are most accurate when they cover signals with similar 
spectra, such as dialogue; the more similar the spectra, the more accurate the measurement. 
This is why we recommend the use of Leq(A) measurements to provision the dialogue level 
value within Dolby E and Dolby Digital bitstreams. The DP570 measurement feature will 
yield best results when the operator measures portions of the program that contain only 
minimal nondialogue material.

Using Infinite Measurements

Infinite measurement mode conforms to the specifications of IEC 60804 (2000‐10). In this 
mode, the calculated value is the long‐term A‐weighted level, Leq(A), integrated over the 
entire period since the measurement was last reset. 

Infinite mode is typically used to measure the entire duration of the program, or several 
long sections of a program, to find the long‐term A‐weighted level. Since Dialogue 
Intelligence is available only in the Dolby LM100 Broadcast Loudness Meter, we suggest 
using the following steps to obtain an estimated Leq(A) value for the entire program:

1.

Review the program, noting the locations of several representative sections where 
dialogue is present.

2. Cue the beginning of the first selected segment.
3. Click Reset to clear past measurements.
4. Measure the first segment, then click Pause to pause measurement.
5. Cue the second segment.
6. Click Pause again to resume measurement.
7. Continue in this fashion, pausing measurement between selected segments until the 

program is completed.

Note: These changes harmonize both the core measurement algorithm and the 

measurement modes between the DP570 and the Dolby LM100 Broadcast 
Loudness Meter. Dialogue Intelligence™, however, is available only in the LM100.