Chapter 5: Using Your PuriCare


Bedding Disposal Operations

• Do not obstruct the front airfoils or rear baffle.

• Large objects should not be placed close

together.

• Keep all materials at least 4 inches inside of the

airfoil, and perform all operations as far to the
rear of the work area as possible.

• Avoid the use of an open flame.

• Avoid using techniques or procedures that

disrupt the airflow patterns of the station.


Final Purging

• Upon completion of work, the station should be

allowed to operate for two to three minutes
undisturbed, to purge airborne contaminants
from the work area.


Wipe-Down

• Wipe down the interior surfaces of the station

with 70% ethanol, or a suitable cleaner, and
allow to dry.

• Inspect and clean the prefilter located under the

HEPA filter.

• Dispose of rubber gloves appropriately, and

have lab coat laundered properly.

• Wash hands and arms thoroughly with

germicidal soap.


Shutdown

• Turn off the fluorescent light and station blower.

Your Next Step


After you understand how to operate and work in the
Bedding Disposal Station, you are ready to proceed to
Chapter 6: Maintaining Your PuriCare.