Speed control adjustment – Labconco PuriCare Bedding Disposal Stations 3840020 User Manual

Page 35

Advertising
background image

31

Product Service 1-800-522-7658

Chapter 6: Maintaining Your PuriCare




Fluorescent Lamp Removal:

1. Remove the front dress panel as described above.
2. Locate the control panel on the front of the cabinet,

which has the filter indicator lights and switches.
Remove the screws located on each lower end of
the panel. Swing the panel upward for access to the
fluorescent lamp, as shown in Figure 6-3.

3. Remove the fluorescent lamp by rotating the lamp

and pulling it straight out of its sockets.

4. Install the new lamp by reversing the removal

procedure.

Speed Control Adjustment


The speed control should only be adjusted by a
qualified certifier.

1. Remove the front dress panel as described

above.

2. Adjust the speed control as required by turning

the screw counterclockwise to increase blower
speed, or clockwise, to decrease the blower
speed, as shown in Figure 6-4.

Figure 6-3