Refer to the following table if your Bedding Disposal
Station fails to operate properly. If the suggested
corrective actions do not solve your problem, contact
Labconco for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Station impellers
and lights won’t
turn on

Unit not plugged into
outlet

Plug the PuriCare into appropriate
electrical service.

Circuit

breakers

tripped

Reset circuit breakers.

Motorized impellers
won’t turn on but
lights work

Impeller wiring is
disconnected

Inspect impeller wiring

Impeller switch is
defective

Replace impeller switch

Motorized impeller is
defective

Replace motorized impeller

Capacitor is defective Replace capacitor

Cabinet impeller
turns on but lights
don’t work

Lamp not installed
correctly

Inspect lamp installation.

Lamp wiring is
disconnected

Inspect lamp wiring