Chapter 6: Maintaining Your PuriCare

3. Remove the two access hole plugs on the top of

the unit.

4. Using the 9/16” deep socket, loosen the filter

clamp nuts located in the access holes. Refer to
Figure 6-6 for further details.

5. With the clamp bolts loosened, the supply filter-

clamping frame should be clear of filter.
Carefully pull the filter straight out of the unit
and discard properly.

6. With the filter removed, inspect the clamping

frame and the frame of the cabinet for damage.

7. Cover the surface of the new HEPA filter gasket

with a light coating of silicone grease, if desired.

8. Install the new HEPA filter by pushing it

straight into the cabinet, ensuring that it is
correctly oriented and the filter fits properly in
the cabinet.

The filter clamp nuts should only be
tightened enough to ensure a proper
seal at maximum tightness, the height
of the filter spring should equal 1"
(25mm). NEVER tighten the clamp
nut beyond this setting. See Figure
6-6.


9. Tighten the clamp nuts uniformly until the filter

gasket is properly compressed against the
cabinet frame. Inspect the seal thoroughly
before proceeding.

10. Reinstall the front dress panel and the access

hole plugs on the top of the unit.

11. Plug the cabinet in and have it recertified before

use.

!