Labconco PuriCare Bedding Disposal Stations 3840020 User Manual

Page 46

Advertising
background image

42

Product Service 1-800-522-7658

Chapter 7: Troubleshooting


PROBLEM

CAUSE

CORRECTIVE ACTION

Cabinet impeller
turns on but lights
don’t work (cont.)

Lamp is defective

Inspect lamp wiring.

Lamp

switch

is

defective

Replace lamp switch

Defective

lamp

ballast

Replace lamp ballast

Service light is on

Blockage of the
exhaust grille

Check the exhaust grille on top of the
unit to ensure they are not blocked or
restricted

Prefilter loading

Replace prefilter

HEPA filter loading

The light will turn on when the filter is
loaded



Advertising