Chapter 7: Troubleshooting
PROBLEM
CAUSE
CORRECTIVE ACTION
Cabinet impeller
turns on but lights
don’t work (cont.)
Lamp is defective
Inspect lamp wiring.
Lamp
switch
is
defective
Replace lamp switch
Defective
lamp
ballast
Replace lamp ballast
Service light is on
Blockage of the
exhaust grille
Check the exhaust grille on top of the
unit to ensure they are not blocked or
restricted
Prefilter loading
Replace prefilter
HEPA filter loading
The light will turn on when the filter is
loaded