Filter indicator system adjustment – Labconco PuriCare Bedding Disposal Stations 3840020 User Manual

Page 36

Advertising
background image

32

Product Service 1-800-522-7658

Chapter 6: Maintaining Your PuriCare


3. Replace the front dress panel and reestablish the

inflow air velocity.



Filter Indicator System
Adjustment

The Filter Indicator System was
adjusted at the factory and should not
require recalibration until the HEPA
filter is replaced.


To adjust system, do the following:

1. Ensure that the prefilter is new and the station is

operating within its airflow tolerance.

2. Remove the two screws that secure the front

dress panel, as shown in Figure 6-3, and remove
the panel.

3. Locate the adjustment screw access plug, on the

left side of the enclosure frame, above the
HEPA filter. Remove the plug. Replace the
dress panel.

4. Using cardboard, or a similar material, block off

approximately ½ of the postfilter. This will
simulate an increased load on the system.

Figure 6-4

!