Routine maintenance schedule, Service operations – Labconco PuriCare Bedding Disposal Stations 3840020 User Manual

Page 32

Advertising
background image

28

Product Service 1-800-522-7658

Chapter 6: Maintaining Your PuriCare

Routine Maintenance
Schedule


After Each Work Session

• Inspect the prefilter and replace if needed.


Weekly

• Using 70% ethanol, or a suitable disinfectant,

surface disinfect the inside of the station, and
the work surface.

• Operate the station blower, noting the indicator

lights.


Monthly (or more often as required)

• Using a damp cloth, clean the exterior surfaces

of the station, particularly the front and top of
the unit, to remove any accumulated dust.

• Replace the prefilter if needed.

• All weekly activities.


Annually

• Have the station recertified by a qualified

certification technician.

• All monthly activities.


Biannually

• Replace the fluorescent lamp.

Service Operations


Prefilter Removal:

The prefilter may be contaminated
with allergenic material. Take
appropriate precautions when
removing and handling the prefilters
to prevent exposure.

!