3 perform single point calibration (set flow) – Casella CEL Apex personal sampling pumps User Manual

Page 14

Advertising
background image

14

TEMP Current internal air temperature.



2.2.2 Manual Reset (Clear Current Sample)

To clear any accumulated sample duration and volume, the instrument must be in Hold Mode.

1.

Press

or

sufficient times to make blink on the display.

2.

Press and hold

.


A CLR message and countdown will be displayed.


3.

Hold the key down until the countdown is complete and CLR

disappears.

On the standard unit, all accumulated values will be cleared.

On the Apex Pro, data from any current event is stored and the

sample (run) terminated ready for a new run to be started.

2.3

Perform Single Point Calibration (Set Flow)

Attach a sampling head and flow tube or other flow-measuring device to the pump’s inlet nozzle to measure
the actual flow as shown in Figures 6 - 8.

TEMP 19.48

2.5

02023

Clamp

Clamping Studs

Apex Pump

Rotameter Float

(Read flow rate
from the top of

the float)

Flowmeter Stand

Rotameter Tube

Sampling Head

Connecting Hose

(5 mm nominal

internal dia.)



TEMP 19.48

2.5

02040

Clamp

Clamping

studs

Stand

Sampling

head

Connecting tube

(5 mm nominal internal dia.)

Flow

Flow

Outlet

Digital Flow meter

Apex Pump

Dry Cal

®

Pr imary Flow Mete r

Figure 6: Calibration with rotameter
Compartment

Figure 7: Calibration with digital calibrator
Compartment