CIRCUTOR DHB Series User Manual

Page 21

Advertising
background image

21

bAud

Baud rate of the RS485 interface

4.8 – 4800 bit/s
9.6 – 9600 bit/s
19.2 – 19200 bit/s
38.4 – 38400 bit/s
57.6 – 57600 bit/s
115.2 – 115200 bit/s

prot

Type of transmission frame of the
RS-485 interface.

r8n2
r8E1
r8o1
r8n1

Addr

Address in the MODBUS network.

The write of the value 0 switches

the interface off

.

0...247

Table 6

SEr

Parameter

symbol

Description

Range of changes

SEt

Write of manufacturer’s settings.

The setting of the value YES

causes the write of standard

parameters into the meter.

The value of manufacturer’s

parameters is presented in the
table 7.

no – do nothing.
YeS – causes the write of

manufacturer’s settings.

SEC

Introduction of a new password.
The introduction of the value 0
switched the password off.

0...60000

HOUR

Setting of the current time. The
introduction of a wrong time can-
cels the introduction of time.
The introduced value will not be
collected.

0.00...23.59

unIt

Backlighting of the unit.

On – unit highlight switched

on.
Off – unit highlight switched

off.

tESt

Display test. The test consists of
a successive lighting up of digital
display segments. Alarm diodes
and unit highlighting diodes
should be lighted .

YeS – causes the test start.
The pressure of the

push-button ends the test.
no – do nothing.