Installation – CIRCUTOR DHB Series User Manual

Page 8

Advertising
background image

8

Fig. 3. Overall dimensions

Fig. 2. Meter fixing

4. INSTALLATION

The meter has separable strips with screw terminals, which enable the
connection of external wires of 2.5 mm

2

cross-section. Strips of input

signals are protected against any accidental disconnection by means
of a screw joint.
One must prepare a hole of 92

+0,6

´ 45

+0,6

mm in the panel, which the

thickness should not exceed 6 mm.
The meter is adapted to be mounted in a panel. The meter must be
introduced from the panel front with disconnected supply voltage. Be-
fore the insertion into the panel, one must check the correct placement
of the seal. After the insertion into the hole, fix the meter by means of
clamps (fig.2).

Advertising