Section 4. Calibration

Voltage Mode Calibration

49

Voltage Mode Calibration

Figure 4.1 Voltage Calibration Setup

Voltage

Program

1. Connect the unit to the digital voltmeter as shown in

Figure 4.1

. See also

“Calibration Setup” on page 48

, step 4.

2. Program the output to 5% of full scale voltage by sending the following

command string from the computer:

VSET2;ISET25 (for 1kW Model 40-25)

3. Adjust voltage programming offset potentiometer R28 (VOFF) until the external

meter reading is within the power supply's voltage program accuracy
specification. Refer to specification tables in

Section 1

.

Example: ±75mV ±(0.3% (5% of 40V)) (for 1kW Model 40-25)

4. Program the output to full scale voltage by sending:

VSET40;ISET25 (for 1kW Model 40-25)

5. Adjust voltage programming full scale potentiometer R51 (VADJ) until the

external meter reading is within the power supply's voltage program accuracy
specification. Refer to specification tables in

Section 1

.

Example: ±75mV ±(0.3% of 40V) (for 1kW Model 40-25)

6. Repeat steps 2 to 5 (between VOFF and VADJ) for best linearity.

V

-

+

VOLTMETER

Power Supply

Negative Output

Power Supply

Positive Output