Section 4. Calibration

Current Mode Calibration

51

Current Mode Calibration

Figure 4.2 Current Calibration Setup

Current

Program

1. Ensure the current shunt has been connected to the power supply as shown in

Figure 4.2

. See also

“Calibration Setup” on page 48

, step 4.

2. Program the output to 5% of full scale current by sending the following

command string from the computer:

VSET40;ISET1.25 (for 1kW Model 40-25)

3. Adjust current programming offset potentiometer R27 (IOFF) until the meter

reading is within the power supply's current program accuracy specification.
Refer to specification tables in

Section 1

.

Example: ±140mA ± (0.15% (5% of 25A)) (for 1kW Model 40-25)

4. Program the output to full scale current by sending:

VSET40;ISET25 (for 1kW Model 40-25)

5. Adjust current programming full scale potentiometer R47 (IADJ) until the meter

reading is within the power supply's current specification. Refer to specification
tables in

Section 1

.

Example: ±140mA ± (0.15% of 25A) (for 1kW Model 40-25)

6. Repeat steps 2 to 5 (between IOFF and IADJ) for best linearity.

V

-

+

VOLTMETER

Power Supply

Negative Output

Power Supply

Positive Output

Current Sensing

Resistor