Section 4. Calibration

Over Voltage Protection (OVP) Calibration

53

Over Voltage Protection (OVP) Calibration

1. Connect the unit as for voltage mode calibration. See

Figure 4.1

. See

“Calibration Setup” on page 48

, step 4. (Optional step.)

2. Program the OVP setting to 50% of the full scale voltage by sending the

following command string from the computer:

OVSET20 (for 1kW Model 40-25)

3. Program the output voltage to about 40% of the OVP set value by sending:

VSET16;ISET25 (for 1kW Model 40-25)

4. Increment the programmed output voltage from the computer until the OVP

fires. The red OVP LED on the front panel will light up.

5. Clear the supply settings by sending:

CLR

6. and reset the OVP condition with:

RST

7. The OVP trip point is in specification if the meter value shown when OVP fires

is within the total of the set trip point ± the OVP program accuracy specification
for the power supply model. See

Section 1

.

Example: 20V ± 400mV (for 1kW Model 40-25)

8. If the OVP trip point was out of specification, adjust potentiometers R19

(OVPREF, coarse adjustment) and R23 (OVPOFF, fine adjustment). Repeat
steps 2 to 5 until the unit operates within specified values again.

9. Check for linearity by testing the OVP trip point at 10% and 90% of full scale

voltage.