Section 4. Calibration
Current Mode Calibration

52

Operating Manual for APG for XT/HPD Series Power Supply

Current

Readback

1. Ensure the current shunt has been connected to the power supply as shown in

Figure 4.2

. See also

“Current Calibration Setup” on page 51

, step 4. Connecting

a DVM is optional.

2. Program the output to 5% of full scale current by sending the following

command string from the computer:

VSET40;ISET1.25 (for 1kW Model 40-25)

3. Perform a readback operation of the output current from the computer:

IOUT?

4. Adjust the current readback offset potentiometer R38 (IZERO) until the

readback value is within the current readback accuracy specification. Refer to
specification tables in

Section 1

.

Example: ±140mA ± (0.15% (5% of 25A)) (for 1kW Model 40-25)

5. Program the output to full scale current by sending:

VSET40;ISET25 (for 1kW Model 40-25)

6. Perform a readback operation of the output current from the computer:

IOUT?

7. Adjust the current readback full scale potentiometer R45 (IGAIN) until the

readback current value is within specification. Refer to specification tables in

Section 1

.

Example: ±140mA ± (0.15% of 25A) (for 1kW Model 40-25)

8. Repeat steps 2 (IZERO) to 7 (IGAIN) for best linearity.