Section 4. Calibration
Voltage Mode Calibration

50

Operating Manual for APG for XT/HPD Series Power Supply

Voltage

Readback

1. Connect the unit to the digital voltmeter (DVM) as shown in

Figure 4.1

.

(Optional step)

2. Program the output to 5% of full scale by sending the following command string

from the computer:

VSET2;ISET25 (for 1kW Model 40-25)

3. Perform a readback operation of the output voltage from the computer:

VOUT?

4. Adjust the voltage readback offset potentiometer R50 (VZERO) until the

readback value is within specification. Refer to the voltage readback accuracy
specification in

Section 1

.

Example: ±75mV ± (0.3% (5% of 40V)) (for 1kW Model 40-25)

5. Program the output to full scale voltage by sending:

VSET40;ISET25 (for 1kW Model 40-25)

6. Adjust the voltage readback full scale potentiometer R49 (VGAIN) until the

readback voltage value is within specification. Refer to the voltage readback
accuracy specification in

Section 1

.

Example: ±75mV ± (0.3% of 40V) (for 1kW Model 40-25)

7. Repeat steps 2 (VZERO) to 5 (VGAIN) for best linearity.