Metro Mobile Computing Workstations MPS-3111 Series Sealed Lead Acid Power Supply User Manual

Page 25

Advertising
background image

Metro

TM

MPS-3111 Series SLA Power Supply Operations Manual Supplement, Doc. No. L01-506

25

InterMetro Industries Corporation

SLA Battery Replacement

Periodically the SLA battery in the MPS-3111 Series SLA Power Supply will no longer hold a
charge and must be replaced with a new battery.

Keep a record, tracking by department, of SLA battery replacement: high use departments will
need more frequent battery replacement than low use departments.

Caution:

For safety, only qualified service personnel should perform the following procedure
to replace the MPS-3111 Series Power Supply battery.

Notice:

New SLA batteries in storage must be maintained by periodically taking them out
and fully charging. Unused SLA batteries in storage will lose charge over time and
if allowed to discharge too completely they will be damaged and unable to hold a
charge. For further information call InterMetro Customer Service.

The following tools are required to replace the MPS-3111 Series Power Supply SLA battery:

• #2

Phillips

screwdriver

• ¼" nut driver (long shaft or shaft extension recommended)
• ¼" box wrench (optional)

To replace the MPS-3111 Series Power Supply SLA battery, perform the following steps:

1. Ensure the ON/Extended Storage switch is set to the “Extended Storage” (OFF) position

and unplug the AC cord.

2. Remove the Power Supply from the workstation base. (See “Removing the MPS-3111

Power Supply” above).