Bolus stopped

If this alarm should occur, it is
very important to check bolus
history to review how much of
the bolus was actually delivered.
Reprogram a bolus with the
amount not delivered, if
required.

Failed batt test

The pump is designed to test
each new battery installed in
the pump. This is a safety fea-
ture to make sure that a battery
that already has low voltage is
not used. If the battery does not
pass the test this alarm will
occur.

Always make sure that you
install a NEW battery into the pump.

This alarm will also happen if a used battery is installed in the
pump or if the installed battery is removed and then put back in.

MiniMed

BOLUS

ESC,ACT to clear

Loose battery cap?

8.35A

STOPPED

Pump dropped or
bumped?
Check bolus history,
Reprogram bolus
if required

BATT TEST

ESC,ACT to clear

Replace with

8.35A

FAILED

Delivery stopped

new battery