22 Chapter 2

Removing the reservoir

1. Remove the infusion set from your body or disconnect the

tubing from the infusion site.

2. Using a counter-clockwise motion, unscrew the tubing

connector from the pump with a 1/2 turn.

3. Pull the reservoir and the connector away from the pump.

4. Dispose of your used reservoir and infusion set in a safe

manner.

Rewinding the pump

WARNING: Always disconnect the infusion set from your body

before inserting or removing the reservoir from the
pump.

1. From the HOME screen, press ACT. Select Prime.

2. Make sure the reservoir is removed from the pump. Press ACT

and Rewind will be selected.

MAIN MENU

Suspend

Basal

MiniMed

Prime

PRIME MENU

Fixed Prime

MiniMed

Rewind

Prime History