Verifying temporary basal delivery – Energizer MMT-511 User Manual

Page 79

Advertising
background image

Basal 67

4. The temporary basal rate will flash. Use the

and

but-

tons to set the desired basal rate. Press ACT.

Verifying temporary basal delivery

The pump will beep/vibrate three times every hour while deliver-
ing temporary basal rate. You can also verify that a temporary
basal rate is delivering by checking the STATUS screen.

1. From the HOME screen, press ESC to access the STATUS

screen. Use the

and

buttons to scroll through the

information. Once you have completed your review, press ESC
to return to the HOME screen.

In the example above the Temp Basal is programmed for 0.6 u/h
for 30 minutes. The time left is 24 minutes, this indicates that 6
minutes have elapsed.

MiniMed

SET EDIT TEMP BASAL

03:30

0.

6

U
H

-

MiniMed

STATUS 7:07A U100u

Temp Basal 0.6 U/H

Duration 0:30

Time left 0:24

Last Bolus N 3.8U

9:07A 11 NOV

Basal 1: 0.8 U/H

Units Remaining: 128U
Battery: Normal

Tue 11NOV2000

S/N# 000000

This shows
Temporary
Basal is active
at 0.6 U/H.

Basal rate that
would have
been delivered
if not for the
Temporary
Basal rate.