84 Chapter 7

Using Auto-off

Provides a safety mechanism that stops insulin delivery after a
defined period of time (1 to 24 hours) if no button presses are
detected. Your pump is programmed at the factory with this feature
turned OFF (0 hours). Discuss with your healthcare professional
what uses and settings are best for you.

1. From the HOME screen, press ACT. Press the

button to

select Utilities. Press ACT.

2. Alarm will be selected. Press ACT.

3. History will be selected. Press the

arrow button to select

Auto-off. Press ACT.

Press the

and

buttons to set the AUTO OFF

DURATION. In this example, the pump will stop delivery of

MiniMed

MAIN MENU

Basal

Prime

Utilities

MiniMed

UTILITIES MENU

Daily Totals

Block

Alarm

MiniMed

ALARM MENU

History

Alert Type

Auto Off