Easy bolus – Energizer MMT-511 User Manual

Page 64

Advertising
background image

52 Chapter 4

Easy bolus

When using vibrate mode, Easy Bolus is limited to 20 steps or
maximum bolus, whichever comes first.

An Easy Bolus is an optional way to deliver a bolus without looking

at your pump. With each press of the

button, you increase

the bolus amount by a set amount, called a “step.” Prior to deliver-
ing an Easy Bolus, you program how many units of insulin will
equal each step. The maximum number of steps depends on the
Easy Bolus step size and the maximum bolus. For example: If the
maximum bolus is set to 15 units and the step size amount is set to
.5, the number of steps is 30.

Example of an Easy Bolus with step amount set at 0.5 units

Once the step amount is set, you are ready to program an Easy

Bolus. From the HOME screen, with each press of the

button,

you will hear a beep or feel a vibration for each step being

Total # Steps = 4

Total # Button Presses = 4

4

3

2

1

0

0

1 Step = 0.5 units

1 Step = 0.5 units

1 Step = 0.5 units

1 Step = 0.5 units

1.0

0.5

1.5

Total # Units

2