64 Chapter 5

Maximum basal rate (max basal)

A maximum basal rate is a safety limit for the amount of basal
insulin that is able to be delivered per hour.

It’s important to discuss your maximum basal rate with your
healthcare professional.

Setting max basal rate

The Max Basal Rate sets the maximum basal rate (units/hour) that
the pump will deliver. This maximum rate will apply to every basal
rate that is set, including a temporary basal rate.

Your pump is programmed at the factory with a maximum basal
setting of two (2.0) units per hour. A maximum basal rate of up to
35.0 units per hour can be set. Discuss with your healthcare profes-
sional what your limit should be. You CANNOT set a maximum
basal rate that is less than any of the currently programmed basal
rates – this includes patterns and temporary basal rates.

1. From the HOME screen, press ACT. Select Basal, then press

ACT.