28 Chapter 2

Manual Prime

1. Make sure that the infusion set is disconnected from your

pump to your body.

2. Hold ACT to Prime. The pump will prepare to prime and will

beep six times.

If you let go of the ACT button, either hold down ACT again to con-
tinue or press ESC to repeat the entire “Rewind” procedure.

3. Continue

to

hold the ACT button until you hear a second set of

six beeps.

WARNING: DO NOT continue until you hear the second set of beeps

and the screen from step 4 appears. Call the 24-hour
Product Help-line for assistance.

MiniMed

MANUAL PRIME

Disconnect from Body

Insert/Lock Reservoir

Hold ACT to Prime

MiniMed

PREPARING TO PRIME

Hold ACT