Default values and ranges, A.1 overview, A.2 most frequently used defaults and ranges – Emerson Process Management MICRO MOTION 2400S User Manual

Page 139: Appendix a default values and ranges, Overview, Most frequently used defaults and ranges, In appendix a

Advertising
background image

Configuration and Use Manual

131

Me

asu

reme

nt P

er
form

anc

e

Defaults

T

rou
b

lesho

o

ti

ng

Compensation

Appendix A

Default Values and Ranges

A.1

Overview

This appendix provides information on the default values for most transmitter parameters. Where
appropriate, valid ranges are also defined.

These default values represent the transmitter configuration after a master reset. Depending on how
the transmitter was ordered, certain values may have been configured at the factory.

A.2

Most frequently used defaults and ranges

The table below contains the default values and ranges for the most frequently used transmitter
settings.

Table A-1

Transmitter default values and ranges

Type

Setting

Default

Range

Comments

Flow

Flow direction

Forward

Flow damping

0.64 sec

0.0–40.96 sec

User-entered value is
corrected to nearest lower
value in list of preset values.
For gas applications,
Micro Motion recommends a
minimum value of 2.56.

Flow calibration factor

1.00005.13

For T-Series sensors,this
value represents the FCF and
FT factors concatenated. See
Section 6.2.2.

Mass flow units

g/s

Mass flow cutoff

0.0 g/s

Recommended setting is 5%
of the sensor’s rated
maximum flowrate.

Volume flow type

Liquid volume

Volume flow units

L/s

Volume flow cutoff

0/0 L/s

0.0–x L/s

x is obtained by multiplying
the flow calibration factor by
0.2, using units of liters per
second.

Meter factors

Mass factor

1.00000

Density factor

1.00000

Volume factor

1.00000