9 configuring the display, 1 update period, 2 language – Emerson Process Management MICRO MOTION 2400S User Manual

Page 75: 3 enabling and disabling display functions, Configuring the display 8.9.1, Update period, Language, Enabling and disabling display functions, Section 8.9, Section 8.9.3 )

Advertising
background image

Configuration and Use Manual

67

Optional Configuration

Requ

ired Co

nfi
g

u

ratio

n

Op

ti

on
al

Co

nfi
g

u

rati

o

n

Using the T

rans

m

itte

r

Us
ing a De

vice

Net T

o

ol

Requ

ired Co

nfi
g

u

ratio

n

Op

ti

on
al

Co

nfi
g

u

rati

o

n

Using the T

rans

m

itte

r

Us
ing a De

vice

Net T

o

ol

Requ

ired Co

nfi
g

u

ratio

n

Op

ti

on
al

Co

nfi
g

u

rati

o

n

Using the T

rans

m

itte

r

Us
ing a De

vice

Net T

o

ol

Requ

ired Co

nfi
g

u

ratio

n

Op

ti

on
al

Co

nfi
g

u

rati

o

n

Using the T

rans

m

itte

r

Us
ing a De

vice

Net T

o

ol

8.9

Configuring the display

If your transmitter has a display, you can configure a variety of parameters that control the display
functionality.

8.9.1

Update period

The Update Period (or Display Rate) parameter controls how often the display is refreshed with
current data. The default is

200 milliseconds

; the range is

100 milliseconds

to

10,000 milliseconds

(10 seconds).

To configure Update Period:

Using ProLink II, see Figure B-3.

Using the display menus, see Figure B-6.

Using a DeviceNet tool, see Table C-9.

8.9.2

Language

The display can be configured to use any of the following languages for data and menus:

English

French

German

Spanish

To set the display language:

Using ProLink II, see Figure B-3.

Using the display menus, see Figure B-6.

Using a DeviceNet tool, see Table C-9.

8.9.3

Enabling and disabling display functions

Table 8-9 lists the display functions and describes their behavior when enabled (shown) or disabled
(hidden).

A116

API: Temperature Outside Standard Range

Info

Yes

No

A117

API: Density Outside Standard Range

Info

Yes

No

A120

ED: Unable to Fit Curve Data

Info

No

No

A121

ED: Extrapolation Alarm

Info

Yes

No

A131

Meter Verification/Outputs at Last Value

Info

Yes

No

A132

Simulation Mode Active

Info

Yes

No

A133

PIC UI EEPROM Error

Info

Yes

No

(1) Can be set to either Informational or Ignore, but cannot be set to Fault.

Table 8-8

Status alarms and severity levels continued

Alarm code

ProLink II message

Default
severity

Configurable?

Affected by
fault timeout?