3 how to characterize, How to characterize – Emerson Process Management MICRO MOTION 2400S User Manual

Page 35

Advertising
background image

Configuration and Use Manual

27

Required Transmitter Configuration

Requ

ired Co

nfi
g

u

ratio

n

Op

ti

on
al

Co

nfi
g

u

rati

o

n

Using the T

rans

m

itte

r

Us
ing a De

vice

Net T

o

ol

When configuring the flow calibration factor:

With ProLink II, enter the concatenated 10-character string exactly as shown, including the
decimal points. For example, using the Flow Cal value from Figure 6-1, enter

19.0005.13

.

With a DeviceNet tool, enter the two factors separately, i.e., enter a 6-character string and a
4-character string. Include the decimal point in both strings. For example, using the Flow Cal
value from Figure 6-1:

-

Enter

19.000

for the flow calibration factor.

-

Enter

5.13

for the temperature coefficient for flow.

6.2.3

How to characterize

To characterize the flowmeter:

1. See the menu flowcharts in Figure 6-2.

2. Ensure that the correct sensor type is configured.

3. Set required parameters, as listed in Table 6-1.

Figure 6-2

Characterizing the flowmeter

Device
· Sensor type

Flow

T Series Config

Straight

tube

Curved

tube

Sensor type?

Density

Flow

Density

ProLink >
Configuration

Sensor type

Flow values

Class: Sensor Information Object (0x67)
Instance: 1
Attribute ID: 3
Data type: USINT
Value:
· 0: Curved tube
· 1: Straight tube
Service: Set

Density values

Class: Calibration Object (0x65)
Instance: 1
Attribute ID 7: K1
Attribute ID 8: K2
Attribute ID 9: FD
Attribute ID 12: D1
Attribute ID 13: D2
Attribute ID 17: DT
Attribute ID 18: FTG
Attribute ID 19: FFQ
Attribute ID 20: DTG
Attribute ID 21: DFQ1
Attribute ID 22: DFQ2
Data type: REAL
Service: Set

Class: Calibration Object (0x65)
Instance: 1
Attribute ID 1: Flow calibration factor
Attribute ID 2: Temperature coefficient for flow
Data type: REAL
Service: Set

DeviceNet tool

ProLink II