Optional Configuration

8.14.2

Configuration procedure

Complete configuration instructions for the enhanced density application are provided in the manual
entitled Micro Motion Enhanced Density Application: Theory, Configuration, and Use. Because of the
complexity of this procedure, Micro Motion recommends using ProLink II for detailed configuration.

If it is necessary to use a DeviceNet tool, refer to the enhanced density manual for application
information, and to the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet:
Device Profile
for complete device profile information.

Basic information on setting up the enhanced density application using a DeviceNet tool is provided
in Figure 8-3.

Volume Conc (SG)
Volume concentration derived from specific
gravity
The percent volume of solute or of material
in suspension in the total solution, derived
from specific gravity

Conc (Dens)
Concentration derived from reference
density
The mass, volume, weight, or number of
moles of solute or of material in
suspension in proportion to the total
solution, derived from reference density

Conc (SG)
Concentration derived from specific gravity
The mass, volume, weight, or number of
moles of solute or of material in
suspension in proportion to the total
solution, derived from specific gravity

Table 8-17

Derived variables and available process variables continued

Available process variables

Derived variable – ProLink II label
and definition

Density at
reference
temperature

Standard
volume
flow rate

Specific
gravity

Concentration

Net
mass
flow rate

Net
volume
flow rate