Emerson Process Management MICRO MOTION 2400S User Manual

Page 34

Advertising
background image

26

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Required Transmitter Configuration

Figure 6-1

Sample calibration tags

Flow calibration values

Two factors are used to define flow calibration:

The flow calibration factor, which is a 6-character string (five numbers and a decimal point)

The temperature coefficient for flow, which is a 4-character string (three numbers and a
decimal point)

These values are concatenated on the sensor tag, but different labels are used for different sensors. As
shown in Figure 6-1:

For T-Series sensors, the value is called the FCF value.

For other sensors, the value is called the Flow Cal value.

Table 6-1

Sensor calibration parameters

Parameter

Sensor type

T-Series

Other

K1

K2

FD

D1

D2

Temp coeff (DT)

(1)

(1) On some sensor tags, shown as TC.

Flowcal

(2)

(2) See the section entitled “Flow calibration values.

FCF

FTG

FFQ

DTG

DFQ1

DFQ2

Other sensors

19.0005.13

0.0010
0.9980

12502.000
14282.000

4.44000

310

12500142864.44

T-Series