Page 66: Configuring cutoffs

Advertising
background image

58

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Optional Configuration

5. Click

Next

.

6. Verify the reference temperature and reference pressure. If these are not appropriate for your

application, click the

Change Reference Conditions

button and enter new values for

reference temperature and reference pressure.

7. Click

Next

. The calculated standard density value is displayed.

If the value is correct, click

Finish

. The value will be written to transmitter configuration.

If the value is not correct, click

Back

and modify input values as required.

Note: The Gas Wizard displays density, temperature, and pressure in the configured units. If required,
you can configure the transmitter to use different units. See Section 6.3.

8.2.2

Using a DeviceNet tool

The Gas Standard Volume Object is used to configure volume flow measurement for gas. See the
flowchart in Figure 8-1.

Figure 8-1

Gas standard volume flow measurement – DeviceNet tool

8.3

Configuring cutoffs

Cutoffs are user-defined values below which the transmitter reports a value of zero for the specified
process variable. Cutoffs can be set for mass flow, liquid volume flow, gas standard volume flow, and
density.

See Table 8-2 for cutoff default values and related information. See Section 8.3.1 for information on
how the cutoffs interact with other transmitter measurements.

Enable gas standard

volume flow

measurement

Set unit

Class: Gas Standard Volume Object (0x64)
Instance: 1
Attribute ID: 7
Data type: BOOL
Value:
• 0 = disabled (and liquid volume flow enabled)
• 1 = enabled (and liquid volume flow disabled)
Service: Set

Set cutoff

(1)

Class: Gas Standard Volume Object (0x64)
Instance: 1
Attribute ID: 8
Data type: REAL
Service: Set

Class: Gas Standard Volume Object (0x64)
Instance: 1
Attribute ID: 5
Data type: UINT
Value: See Table 6-4
Service: Set

Set reference density

of gas

(2)

Class: Gas Standard Volume Object (0x64)
Instance: 1
Attribute ID: 4
Data type: REAL
Service: Set

(1) See Section 8.3.
(2) The Gas Wizard is provided only

with ProLink II. If you are not
using ProLink II, you must
supply the required reference
density.