Step – Epson Weighing Indicator XK3190-A1+P User Manual

Page 14

Advertising
background image

X K 3 1 9 0 - A 1

+P

communication mode parameters.

(2) The sequence of parameters setting as following:

Connecting the loadcell , the indicator enters into working mode which is shown in

the Graph 2-3.

Insert the setting jumper into 15-pin socket at the back of indicator. (There is a 15-pin

plug in the packing carton with its 14-pin shorte connected with 15-pin. ).

Setup communication parameters as following:

Step Operation Display

Explanation

1

Press[CALB]

Plug in calibration jumper

2

Press[Enter]

[E **]

Don’t change this data and enter next step.

3

Press[Enter]

[dc **]

Don’t change this data and enter next step.

4

Press[Enter]

[Pon **]

Don’t change this data and enter next step.

5

Press[Enter]

[F

******]

Don’t change this data and enter next step.

6

Press[Enter]

Press[Enter]

[H *****]

[L *****]

[td **.**]

Don’t change this data and enter next step.

7

Press[1]

Press[Enter]

[A d r **]

[A d r01 ]

Communication address (01~26)

Example: 1

8

Press[1]

Press[Enter]

[bt *]

[bt 1]

Baud rate of serial communication (0~4)

which represent baud rate of

600,1200,2400,4800,9600

9

Press[0]

Press[Enter]

[tF *]

[tF 0]

Serial communication mode:

0- Continous transmitting mode,

unreceiving

1- command responding mode

10

Weighing

mode

End of communication parameters setting

( Please note the explanation. Don’t change other parameters setting carelessly )

4.7 Date, Time Set-up

(1) Under weighing status of the indicator, press [ Date ] key, and date sign lamp is

lighted. Indicator display original data at this time. If it is correct, then press