3. Calibration

According to the Graph 2-3, connecting Indicator and loadcell, and the indicator enters

weighing mode.

1. Connecting calibration jumper to the 15- pin socket on the indicator back panel.

(There is a 15-pin plug in the packing carton with its 14-pin short connected with

15-pin. )

2. According to table 4-3, carry out calibration step by step.

Table 4 - 1

step Operation

Display

Explanation

1 press [ CALB]

after calibration jumper is inserted

2

press [1 ] [0 ]

press [Enter ]

[E ** ]

[E 10 ]

Enter division value selecting

1/2/5/10/20/50/100/200

Example: 10

3

press [ 0 ]

press [ Enter ]

[ dc * ]

[ dc 0 ]

Enter decimal point (0 -4)

Example: without Decimal point 0

4

press [ 2 ] [ 3 ]

press [ Enter ]

[pon XY ]

[ pon 23 ]

Enter Zero range

X: zero set range (1-5) it stands for

2% 4% 10% 20% 100% of F.S

Y: power on Zero set range it

Stands for (1-5)

2% 4% 10% 20% 100% of F.S

example 23

5

press [ 3 ] [ 0 ]

[ 0 ] [ 9 ] [ 0 ]

[F ***** ]

[ F 30090 ]

Enter overload warning value when

inputting F value,and re-pressing

[ Enter ] key, the calibration will begin.

directly pressing [ Enter ],will enter into

the tenth step. If pressing [ weighting ],

Indicator will return to weighing status.

6 press [ Enter ]

Example: 30090

7

[ noLoAdn] Confirming Zero position. At this time

there is no load on the scale.Pressing

[Input] when the stable light is on.