3 connecting loadcell to indicator – Epson Weighing Indicator XK3190-A1+P User Manual

Page 6

Advertising
background image

X K 3 1 9 0 - A 1

+P

15

In calibration or parameters setting mode, pressing [weigh], it will exit from original
calibration or setting mode and return to weighing mode.

2.3 Connecting Loadcell to Indicator

1. The 9-pin socket is used for the link-up of loadcell, which has been clearly shown

in the graph 2-3.

2. The 4-pin shielded cable is used, +S must be short connected with +E, -S and -E. The

indicator does not have the function of long distance compensation.

3.

▲ !Indicator must be reliably connect ed to Loadcell and shielded-cable of

loadcell must be reliably connect ed to underground . If indicator is

powered on, the user should not insert or withdraw the plug in order to

protect the indicator and loadcell.

4.

▲ !Sensor and indicator are static sensitive devices; you must adopt anti-static

measures. In order to protect the operator ,indicator, and relevant devices,
you should install lightning rod in the thunderstorm frenquently
happening area.


1

2

3

4

5

6

7

8

9

-E

-S

Shield

+E

+S

-IN

+IN











5 Shield

8 -IN

9 +IN

1 -E

2 -S

7 +S

6 +E

+

_

Terminal of Indicator

Terminal of Sensor

Bridge of Power Supply

Signal Output










(Graph 3-2 ) Connection of the load cell