Remote control via lan (option), Remote control via lan (option) -5 – RIGOL DP832 User Manual

Page 127

Advertising
background image

Chapter 3 Remote Control

RIGOL

DP800 User’s Guide

3-5

Remote Control via LAN (option)

1. Connect the device

Connect the power supply to your PC or the LAN of your PC using a network

cable.

2. Configure network parameters

1) DHCP mode:

If the network supports DHCP, the DHCP server in the network assigns

network parameters (IP Address, Subnet Mask, Gateway and DNS) for the

power supply automatically.

2) Auto IP mode:

When the network doesn’t support DHCP, the DHCP mode of the power

supply is disabled or the power supply is connected to the PC directly, the

power supply selects Auto IP mode and acquires the IP address from

169.254.0.1 to 169.254.255.254 and subnet mask 255.255.0.0

automatically.

3) Manual IP mode:

Enable the Manual IP mode and disable the DHCP mode as well as the Auto

IP mode to set the network parameters manually.

If the power supply is connected to the PC directly, set the IP Addresses,

Subnet Masks and Gateways for both of the PC and the power supply. The

Subnet Masks and Gateways of the power supply and PC must be the same

and the IP Addresses of them must be within the same network segment

(for details, refer to the TCP/IP protocal). An example is shown in the table

below.

Network Parameters

PC

Power Supply

IP Address

Subnet Mask

Default Gateway

192.16.3.3

255.255.255.0

192.16.3.1

192.16.3.8

255.255.255.0

192.16.3.1

If your power supply is connected to the LAN of your PC, please acquire a

set of valid network parameters (IP Address, Subnet Mask, Gateway and

DNS) from your network administrator and configure the network

parameters of the power supply according to the description in "To Set the

LAN Parameters".

Advertising