Sense Working Mode

When the output current of the power supply is high, the voltage drop on the load

lead cannot be ignored. To ensure that the load can acquire the correct voltage drop,

CH2 of DP821 and DP811 (both provide 10A output current) provide the Sense

(remote sense) working mode. In this mode, the instrument detects the voltage at

the load terminal instead of the voltage at the output terminal of the power supply;

in this way, the instrument can automatically compensate for the voltage drop

caused by the load lead to ensure that the power supply output value set by users is

consistent with the voltage acquired by the load.

The Sense connecting method at the front panel is as shown in the figure below.

Operation Procedures:

1. Connect the channel output terminals and Sense terminals at the front panel of

the instrument to the two ends of the load respectively, as shown in the figure

above. Pay attention to the polarity when making connections.

2. For DP821: after you select CH2 (8V/10A), press Sense to turn on the Sense

function.

For DP811: after you select the desired range, press Sense turn on the Sense

function.

In addition, the output interface at the rear panel of DP811 can also be used for the

Tip:
 When the output current is high, to acquire the best output characteristics,

the load leads should be as short as possible and it is recommended that you

twist the load leads together.

 Twisted pairs are recommended for the Sense leads and do not twist the

Sense leads and load leads together.

Lead Equivalent
Resistance

Sense Lead

Load Lead

Sense Lead

Load Lead

Lead Equivalent
Resistance