To Set the Number of Output Groups

The number of output groups is defined as the number of groups of preset voltage

and current values that the power supply outputs in each cycle. Press Groups and

use the numeric keyboard or the left/right direction key and knob to input the desired

value. The range is from 1 to 2048.

To Set the Number of Cycles

The number of cycles is defined as the number of times that the instrument performs

timing output according to the preset voltage and current. Press Cycles to set the

number of cycles to "Infinite" or use the numeric keyboard or the left/right direction

key and knob to input the desired value. The range is from 1 to 99999.

To Set the End State

The end state refers to the state of the instrument after it finishes outputting the

total number of groups of voltage and current values when the number of cycles is a

specified value. Press End State to select "Outp Off" or "Last".
 Outp Off: the instrument turns off the output automatically after finishing the

output.

 Last: the instrument stops at the output state of the last group after finishing

the output.

Note: When the number of cycles is set to "Infinite", the end state is invalid.

To Edit the Timer Parameters

You can edit the timer parameters manually or using the built-in templet.

Tip
 The total number of groups in timing output = the number of output groups*

the number of cycles

 The power supply will terminate the timer function when the total number of

groups of outputs is finished. At this point, the state of the power supply

depends on the setting in "To Set the End State".