RIGOL DP832 User Manual

Page 70

Advertising
background image

RIGOL

Chapter 2 Front Panel Operations

2-20

DP800 User’s Guide

Ramp

The Ramp waveform is as shown in the figure below. The instrument

determines the amplitude of the Ramp according to the maximum and

minimum currently set, determines the period according to the total

number of points (denoted by P) and the time interval currently set and

determines the Ramp waveform according to the symmetry (denoted by

Sym) currently set. When creating parameters, the instrument draws

int(P*Sym)

[1]

values from the rising edge of the preset Ramp waveform at

the same time interval and draws P-int(P*Sym) values from the falling

edge of the preset Ramp waveform at the same time interval. The timing

time is determined by the time interval currently set.

T

t

Symmetry=t/T*100%

Note

[1]

: int(P*Sym) refers to rounding P*Sym (discard the decimal part).

Stair Up

The Stair Up waveform is as shown in the figure below. The instrument

determines the Stair Up waveform according to the maximum (denoted by

MAX), minimum (denoted by MIN), total number of points (denoted by P,

P≥10) and time interval currently set and creates P parameters from MIN

to MAX at the step of (MAX-MIN)/(P-1). The timing time is determined

by the time interval currently set.

Stair Up

Stair Dn

The Stair Dn waveform is as shown in the figure below. The instrument

determines the Stair Dn waveform according to the maximum (denoted by

MAX), minimum (denoted by MIN), total number of points (denoted by P)

and time interval currently set and creates P parameters from MIN to MAX

at the step of (MAX-MIN)/(P-1). The timing time is determined by the

time interval currently set.

Stair Dn

Advertising