RIGOL DSG3000 Series User Manual

Page 57

Advertising
background image

RIGOL

DSG3000 Quick Guide

7

9. Power Key

Turn on or off the signal generator. When this key is

turned off, the signal generator is in standby state and it

will be powered off only when the power switch at the

rear panel is turned off and the power cord is removed.


Press System Power Status to select “Open” or

“Default”. When “Default” is selected, after the

instrument is powered on and the power switch at the

rear panel is turned on, you need to press this key to

start the instrument. When “Open” is selected, the

instrument will start automatically after it is powered on

and the power switch at the rear panel is turned on.

10. Built-in Help System

To get the help information of any front panel key or

menu softkey, press this key and then press the desired

key.

11. View Switch Key

It is used to switch the interface display mode to normal

mode or parameter zoom-in mode.

12. Trigger Control Key

When the trigger type of Sweep is “Key”, press this key
once to trigger a sweep.
When the trigger mode of Pulse is “Key”, press this key
once to enable a pulse modulation.

13. External Modulation Input Connector

When the modulation source of AM, FM and

Φ

M is

external, this connector is used to input the external

modulating signal.