RIGOL DSG3000 Series User Manual

Page 80

Advertising
background image

RIGOL

30

DSG3000 Quick Guide

Make sure that at least two points are set in the step sweep or list

sweep.

(2) The dwell time of list sweep is incorrect

Press SWEEPList Swp to enter the sweep list editing interface.

Make sure whether the dwell time of the list sweep is correct.

If the dwell time is incorrect, re-edit it; if it is correct, turn to the next

step.

Press SWEEPSweep Type to make sure that “List” sweep is
selected.

If “Step” sweep is currently selected, the signal generator sweeps

according to the dwell time of the step sweep.

(3) The list sweep information recalled from the register is lost

The list sweep information cannot be stored in the status register of

the instrument as a part of the instrument state.

The signal generator can only use the current list for sweep and you

can store the list sweep data to the local directory of the instrument.

(4) In list or step sweep, the amplitude does not change

Make sure that the sweep manner is set to “Level” or “Freq&Level”.

If the sweep manner is currently set to “Freq”, the amplitude will not

change.

6. The USB storage device cannot be recognized:

(1) Check whether the USB storage device can work normally when connected

to other instrument or PC.

(2) Make sure that the USB storage device used is Flash type. This instrument

does not support hardware USB storage device.

(3) Restart the instrument and insert the USB storage again to check whether it

can work normally.

(4) If the USB storage device still cannot be used normally, please contact

RIGOL.


7. No response or incorrect response when key is pressed:

(1) After starting the instrument, make sure that whether all the keys are

irresponsive.

(2) Press SystemSelf-testKey Test to enter the keyboard test

interface. Press the function keys at the front panel sequentially and check

whether some key gives no response or incorrect response when it is

pressed.

(3) If the above failures occur, the keyboard connecting wire might be loose or

the keyboard is damaged. Please do not disassemble the instrument and

contact

RIGOL.

8. Performance specification test is not passed:

(1) Check whether the RF signal generator is within calibration period (1 year).