8

DSG3000 Quick Guide

14. LF Output Connector

When the backlight of LF goes on, this connector is used
to output LF signal.

15. RF Output Connector

When the backlight of RF goes on, this connector is used
to output RF signal and RF sweep signal.

When the backlights of RF and MOD go on, this
connector is used to output RF modulated signal.

CAUTION

To avoid damage to the instrument, the reverse DC voltage on

the RF output connector cannot exceed 50 V, the reverse power

must be less than +10 dBm in 9 kHz~1 MHz frequency range

and +30 dBm in 1 MHz~6 GHz frequency range.