RIGOL DSG3000 Series User Manual

Page 79

Advertising
background image

RIGOL

DSG3000 Quick Guide

29

4. The setting is correct but the waveform output is incorrect:

(1) There is no RF output

Check whether the signal connecting wire is tightly connected to the

corresponding [RF OUTPUT 50Ω] connector.

Check whether the connecting wire has internal damage.

Check whether the connecting wire is tightly connected to the test

instrument.

Check whether the backlight of RF goes on. If not, press this key to
illuminate the backlight. At this point, the RF label is displayed in the

status bar of the user interface and the RF output is enabled.

Check whether the output amplitude of the RF signal is too small.

Adjust the output amplitude properly.

(2) The RF output is not modulated

Check whether the signal connecting wire is tightly connected to the

corresponding [RF OUTPUT 50Ω] connector.

Check whether the connecting wire has internal damage.

Check whether the connecting wire is tightly connected to the test

instrument.

Check whether the backlights of MOD and RF are both illuminated
and check whether the modulation Switch is set to “On”.

Check whether the modulation parameter setting is correct and adjust

the parameter appropriately.

If external modulation source is used, make sure that the external

source is correctly connected and outputs signal. At the same time, the

external source should work within the specified range of the signal

generator.

5. The sweep is abnormal:

(1) The seep stagnates

The sweep progress bar is displayed in the frequency area/amplitude area

in the user interface indicating that the sweep is in progress. If the sweep

stagnates, please check:

At least one manner of sweep is enabled. Press SWEEPSweep to
select “Freq”, “Level” or “Freq&Level”.

If single sweep mode is selected, press Single and the instrument

performs a sweep if the trigger condition is met.

If the sweep trigger mode is not set to auto, press SWEEPTrig
“Auto” to make sure whether the sweep stagnates due to the loss of

sweep trigger.

If the point trigger mode is not set to auto, press SWEEPPoint

Trig  “Auto” to make sure whether the sweep stagnates due to the

loss of point trigger.

Make sure whether the dwell time is too large or too small that the

sweep cannot be seen.