5. INITIAL SETTINGS

14

DATA 2 output setting

1) Press [MENU] [9] and [4]. The DATA 2 OUTPUT SETUP menu appears.

183/V2.0

Talker ID

GP

LC

DE

Output Data (00-90 sec)80%

DATA2 OUTPUT SETUP

MENU : Escape

ENT : Enter

NMEA Fmt.

183/V1.5

: Select

Setting shown here are default settings.

1. AAM:00

APA:00

APB:04 BOD:00

2. BWC:00 BWR:00 BWW:00 GGA:00

3. GLL:01

RMB:01 RMC:01 VTG:01

4. WCV:00 WNC:00 WNR:00 WPL:10

5. XTE:00

ZDA:01

GNS:00 ZTG:00

Figure 5-2 DATA 2 OUTPUT SETUP menu


2) Follow the procedure for setting DATA 1 output.

DATA 3 output setting

The DATA 3 can output NMEA 0183 (V1.5/V2.0) /IEC 61162-1 data or log pulse by
selecting inner jumper wire. For NMEA 0183 (V1.5/V2.0) /IEC 61162-1, the same signal of
DATA 1 is output from DATA 3.

Selection of NMEA0183 or log pulse

Output data

NAV board

NMEA 0183 (V1.5/

V2.0)/IEC 61162-1

Log pulse

JP4 (2 pcs.)

ON

OFF

JP5 (3 pcs.)

OFF

ON


Rate of log pulse output

500 mA Max.
50 VDC


Procedure for setting of log pulse rate

1) Press [MENU] [9] and [3].
2) Press

!

or

"

to select DATA 3. Log Pulse.

3) Press

#

or

$

to select log pulse for external equipment; 200ppm or 400ppm.

4) Press the [ENT] key.
5) Press the [ESC] key.