For Compass Units with the SHARK Controller

Flash codes indicate the nature of an abnormal condition directly from the
SHARK Information Gauge. Without the use of any servicing tools, the condition
can be simply diagnosed.

Flash
Code

Description

1

User Fault

Possible stall timeout or user error. Release
the joystick to neutral and try again.

2

Battery Fault

Check the batteries and cabling. Try charging
the batteries. Batteries may require replacing.

3

Right Motor Fault

Check the right motor, connections and

cabling.

4

Left motor fault

Check the left motor, connections and

cabling.

5

Right park brake fault

Check the right park brake, connections and

cabling.

6

Left park brake fault

Check the left park brake, connections and
cabling.

7

SHARK Control Unit Fault Check the SHARK Communications Bus con-

nections and wiring. Replace the Control Unit

8

SHARK Power Module
Fault

Check SHARK connections and wiring.
Replace the Power Module.

9

SHARK Communications
Fault

Check SHARK connections and wiring.
Replace the SHARK Control Unit.

10 Unknown Fault

Check all connections and wiring. Consult a
service agent.



The SHARK modules are not user serviceable. Do not attempt to repair Shark

modules.

Note that joystick OONAPU (Out Of Neutral At Power Up) is not a fault. Simply by
removing your hand from the joystick and allowing it to return to the neutral position,
the fault will immediately clear. If the condition persists after removing your hand, the
joystick may be damaged. Consult a Service agent.