Operation/diagnostics, For compass units with the a-series controller – Golden Technologies GP601 SS User Manual

Page 26

Advertising
background image

Compass Onwer’s Manual – Models GP600 SS, GP600 CC, GP601SS, GP601 CC Revised 060208

26

OPERATION/DIAGNOSTICS

For Compass Units with the A-SERIES Controller

Flash codes indicate the nature of an abnormal condition directly from the A – Series
Information Gauge. Without the use of any servicing tools, the condition can be
simply diagnosed.

Flash

Code

Description

Recommended Action

1

User Fault

Possible stall timeout or user error. Release

the joystick to neutral and try again.

2

Battery Fault

Check the batteries and cabling. Try charging

the batteries. Batteries may require replacing.

3

Right Motor Fault

Check the right motor, connections and
cabling.

4

Left motor fault

Check the left motor, connections and
cabling.

5

Right park brake fault

Check the right park brake, connections and
cabling.

6

Left park brake fault

Check the left park brake, connections and

cabling.

7

User Interface Fault

Display PCB Fault
Consult a Service Agent

8

Controller Fault

Consult a Service Agent

9

A – Series Bus

Communications Fault

Consult a Service Agent












Advertising